IEC 61010-2-081:2019 pdf free download.Safety requirements for electrical equipment for measurement, control and laboratory use – Part 2-081: Particular requirements for automatic and semi-automatic laboratory equipment for analysis and other purposes.
IEC 61010-2-081:2019 applies to automatic and semi-automatic laboratory equipment for analysis and other purposes. Automatic and semi-automatic laboratory equipment consists of instruments or systems for measuring or modifying one or more characteristics or parameters of samples, performing the complete process or parts of the process without manual intervention. Equipment forming part of such a system is within the scope of this document.
5.4.101 Removal of equipment from use for repair or disposal Instructions shall be provided to the RESPONSIBLE BODY for eliminating or reducing HAZARDS involved in removal from use, transportation or disposal,or appropriate contact information shall be provided in the instructions.
7 Protection against mechanical HAZARDS
This clause of Part 1 is applicable except as follows:
7.3.2 Exceptions
Replacement:
Replace the text in item b)3)by the following new text:
there are warning markings prohibiting access by untrained OPERATORS. Markings shall be placed within the area requiring maintenance where they can alert the OPERATOR to the HAZARD.As an alternative, symbol 14 of Table 1 can be used, with the warnings incuded in documentation.
10 Equipment temperature limits and resistance to heat
This clause of Part 1 is applicable.
11 Protection against HAzARDS from fluids and solid foreign obiects
This clause of Part 1 is applicable.
12 Protection against radiation, including laser sources, and against sonic and ultrasonic pressure
This clause of Part 1 is applicable.
13 Protection against liberated gases and substances, explosion and implosion

                                           Related Information                                             Download
PS:Thank you for your support!
AS NZS 60947.5.9:2015 pdf download AS Standards

AS NZS 60947.5.9:2015 pdf download

AS NZS IEC 60947.5.9:2015 pdf download.Low-voltage switchgear and controlgear Part 5.9: Control circuit devices and switching elements——Flow rate switches. 8.3.3.2.2 Time delay before availability The test is performed with the flow rate switch connected to a test...
Read More
AS 60990:2018 pdf download AS Standards

AS 60990:2018 pdf download

AS IEC 60990:2018 pdf download.Methods of measurement of touch current and protective conductor current. 4.2 Test transformer The use of a test transformer for isolation is optional. For maximum safety, a test transformer for isolation (T2 in...
Read More
IEC 60674-1:1980 IEC Standards

IEC 60674-1:1980

IEC 60674-1:1980 pdf download.Specification for plastic films for electrical purposes - Part 1: Definitions and general requirements. INTRODUCTION This standard is one of a series which deals with plastic films for electrical purposes. The series will consist...
Read More

LEAVE A REPLY

Anonymous netizen Fill in information