ANSI N42.14-1999 pdf download American National Standard for Calibration and Use of Germanium Spectrometers for the Measurement of Gamma-Ray Emission Rates of Radionuclides
3. Definitions
The word “shall” denotes a requirement, the word “should” denotes a recommendation, and the word “may” denotes permission. To conform to this standard, Ge spectrometry shall be performed in accordance with the requirements of this standard, but not necessarily with its recommendations; however, justification should be documented for deviations from a recommendation.
For the purposes of this standard, the following terms and definitions apply. IEEE Std 100-1996 should be referenced for terms not defined in this clause.
3.1 abundance: See: emission probability per decay [ P ( E ) or P γ ( E )].
3.2 activity (A): The expected number of spontaneous nuclear decays (transformations) in unit time from a specified energy state (excluding prompt decays from a lower nuclear level) for a given amount of a radionu- clide. Its standard unit (SI) is the becquerel (Bq), where one Bq equals one decay per second. Activity has often been expressed in curies (Ci), where 3.7 × 10 10 Bq equals 1 Ci, exactly.
3.3 ADC conversion gain: The number of channels over which the full amplitude span can be spread; usu- ally 2048–8192 channels are used for Ge gamma-ray spectrometry.
3.4 analog-to-digital converter (ADC): An electronic device used to convert the amplitude of a voltage pulse from analog to digital format.
3.5 attenuation: The net loss at the detector of primary photons of a given energy resulting from their inter- action with matter either due to the occurrence of scattering or absorption in the sample or in material between the sample and the detector crystal.
3.6 background: Spectral data including peaks not caused by the source but rather resulting from radioac- tive decay occurring in the surrounding environment or resulting from cosmic-ray interactions in or adjacent to the detector. See also: continuum.
3.7 baseline: See: continuum.
3.8 calibration: The determination of a value that converts a measured number into a desired physical quan- tity (e.g., pulse height into photon energy, or counts per second into emission rate).
3.9 cascade summing: See: coincidence summing of x and gamma rays.
3.10 cascade transitions: Gamma rays in the radioactive decay of a single atom that are emitted sequen- tially and within the resolving time of the spectrometer.
3.11 coincidence summing of x and gamma rays: The simultaneous detection of two or more photons originating from a single nuclear disintegration that results in only one observed (summed) pulse. Syn: cascade summing; true coincidence summing.

                                           Related Information                                             Download
PS:Thank you for your support!
ANSI AWS A5.03-1999(R2007) pdf download ANSI Standards

ANSI AWS A5.03-1999(R2007) pdf download

All standards (codes, specifications, recommended practices, methods, classifications, and guides) of the American Welding Society (AWS) are voluntary consensus standards that have been developed in accordance with the rules of the American National Standards Institute (ANSI). When...
Read More
ANSI AWS A5.20-1995 pdf download ANSI Standards

ANSI AWS A5.20-1995 pdf download

Note: The primary purpose of AWS is to serve and benefit its members. To this end, AWS provides a forum for the exchange, consideration, and discussion of ideas and proposals that are relevant to the welding industry...
Read More

LEAVE A REPLY

Anonymous netizen Fill in information